Assessment of the efficacy of an indigeniously developed pulse oximeter dental sensor holder for pulp vitality testing. |
|
V Gopi Krishna, D Kandaswamy, T Gupta
Department of Conservative Dentistry and Endodontics, Meenakshi Ammal Dental College, Chennai, India
Click here for correspondence address and email
|
|
 |
|
Abstract | | |
Current routine methods of assessment of pulp vitality rely on stimulation of nerve fibres and give no direct indication of blood flow within the pulp. Pulse oximeter is a proven, atraumatic method of measuring vascular status, by evaluating oxygen saturation. This study explores the use of customized dental pulse oximeter sensor holder to assess pulpal vascular oxygen saturation in permanent teeth. Pulse oximeter readily differentiated between known vital and nonvital teeth. Vital teeth consistently provided oxygen saturation values that were lower than the values recorded on the patient's fingers. Pulse oximeter is an accurate, atraumatic clinical alternative to the present electrical and thermal methods of assessing pulp vitality in teeth. Keywords: pulse oximeter, pulp vitality, sensor holder
How to cite this article: Gopi Krishna V, Kandaswamy D, Gupta T. Assessment of the efficacy of an indigeniously developed pulse oximeter dental sensor holder for pulp vitality testing. Indian J Dent Res 2006;17:111 |
How to cite this URL: Gopi Krishna V, Kandaswamy D, Gupta T. Assessment of the efficacy of an indigeniously developed pulse oximeter dental sensor holder for pulp vitality testing. Indian J Dent Res [serial online] 2006 [cited 2023 Mar 22];17:111. Available from: https://www.ijdr.in/text.asp?2006/17/3/111/29880 |
Introduction | |  |
The assessment of pulp vitality is a crucial diagnostic procedure in the practice of endodontics. Conventionally, the dentist has relied on tests that depend on the patient's perceived response to a stimulus as well as the dentist's interpretation of that response. These methods include thermal stimulation (as in the case ofheat or cold application), electric stimulation or direct dentin stimulation (test cavity).
These tests fall short of the ideal pulp vitality testing on several criteria. Each of the methods described, tests only sensory nerve fibre response. Stimulation of nerve fibres is not the ideal method to determine vitality status. Several authors [1],[2],[3],[4] have stated that vascular supply and not innervation is the most accurate determinant for assessing pulp vitality. As a result, teeth that have temporarily or permanently lost their sensory function (e.g. teeth damaged by trauma. or teeth that have undergone orthognathic surgery) will be non responsive to these tests. However, they may have intact vasculatures [5],[6]. Moreover, the nervous tissue, being highly resistant to inflammation, may remain reactive long after the surrounding tissues have degenerated. Therefore, thermal and electric tests may give false positive response if only the pulp vasculature is damaged [2]. Further, all these tests have the potential to produce an unpleasant and occasionally painful sensation and inaccurateresults may be obtained [3].
Recent attempts to develop a method for determination of pulpal circulation have involved the use of laser doppler flowmetry dual wavelength spectrophotometry and pulse oximetry. Although laser doppler flowmeter has met with some success in medical applications, its use in dentistry has been hampered by the sizeable expense, tack of reproducibility and sensitivity of the device to motion. Dual wavelength spectrophotometry has been examined only in the laboratory setting so far and only detects quantitatively the presence of haemoglobin and not the circulation ofblood [7].
Pulse oximeter is a non-invasive oxygen saturation monitoring device widely used in medical practice for recording blood oxygen saturation levels during the administration of intravenous anaesthesia through the use of finger, foot or ear probes. It was invented by Takuo Aoyagi, a biomedical engineer working for the Shimadm Corporation in Kyoto, Japan, in early 1970's [8]. Pulse oximeter uses red and infrared wavelengths to transilluminate a tissue bed, detecting absorbance peaks due to pulsatle blood circulation and uses this information to calculate oxygen saturation and pulse rate [3]. It is the most commonly employed technique to measure oxygen saturation because of its ease and economy.
Different researchers have arrived at different conclusions regarding the efficacy of using pulse oximetry to diagnose pulp vitality. Schnettler and Wallace [9] reported a correlation between pulp and systemic oxygen satiation readings using a modified ear pulse oximeter probe on a tooth. Thus, they recommended its use as a definitive pulp vitality tester. Kahan et al [10] designed, built and tested a reflectance tooth probe using Biox 3740 oximeter. Pulse wave readings from the teeth were found to be synchronous with the finger probe but not consistently. They concluded that the accuracy of the commercial instrument was disappointing and in its present form, it was not considered to have predictable diagnostic value.
The critical requirement of using pulse oximeter in dentistry is that
P. The sensors should confirm to the size, shape and anatomical contours of teeth.
The light emitting diode sensor and the photoreceptor should be parallel to each other so that all the light emitted by light emitting diode sensor is received by phototoreceptor sensor. P~ Thepulseoximeterprobe should beheld firmly onto the tooth to ensLire accurate measurements.
Taking into consideration, the above requirements, the aim of our study was
To design and built a customized pulse oximeter dental sensor holder and evaluate its efficacy in determining the vitality ofpermanentteeth.
P. To determine the values of oxygen saturation in normal, permanent maxillary incisors and canines using pulse oximeter.
To evaluate the relationship between tooth pulse oximeter readings and systemic oxygen saturation measured on the index finger.
Materials and methods | |  |
The sample population consisted of one hundred patients aged 15 to 40 years with normal maxillary central incisors, lateral incisors and canines. Sample criteria required the teeth to be free of caries, fracture or discolouration, with healthy periodontal status, no radiographic periapical changes and complete formation of root apex. The patient's dental history had to be negative for any history oftraumato face, mouth or teeth.
As a control population to confirm the pulse oximeter's evaluation of pulp vascularity, thirty nonvital primary teeth with complete endodontic fillings were tested. Pulse oximeter readings from the patient's fingers served as the control sample for comparison of pulp oxygen saturation values with the patient's systemic oxygen saturation values.
A Nellcor OxiMax' Dura Y D-YS multisite oxygen sensor was used to record oxygen saturation values on teeth. It was held on the tooth using a customized sensor holder.The same sensor with sensor holder used on tooth was used to measure finger oxygen saturation values. Testing equipment consisted of the Nellcor OxiMax N550 pulse oximeter (Tyco Healthcare Group LP, U. SA).
Procedure
The patient's oxygen saturation values was first measured on the index finger using a finger probe. The patient's central incisors, lateral incisors and canines were then evaluated by placing the customized sensor holder with the sensor onto the tooth as shown in [Figure - 1]. The customized sensor holder allowed the light emitting diode sensor and the photoreceptor to be parallel to each other so that all the light emitted by light emitting diode sensor is received by phototoreceptor sensor. The values were recorded after 30 seconds of monitoring each tooth. Oxygen saturation results were recorded and analysed using the Pearson correlation analysis to evaluate the relationship between the finger and tooth pulse oximeter readings [Table - 1].
Results | |  |
The thirty non-vital teeth (control) recorded oxygen saturation values of 0%. The mean oxygen saturation value for maxillary central incisor was 79.31 (SD=3.07), for maxillary lateral incisors 79.61 (SD=2.73) and for maxillary canines 79.85 (SD=2.09). Their control values, measured on the patient's index fingers, averaged 97.58% (SD=0.57). Pearson's correlation analysis of each individual's finger and tooth oxygen saturation values showed a 0.15 correlation for maxillary central incisors, 0.22 for maxillary lateral incisors and 0.13 for maxillary canines.
Discussion | |  |
This study confirmed the ability of the customized pulse oximeter dental probe to differentiate between vital and non-vital teeth. All vital teeth provided consistent oxygen saturation readings and all nonvital teeth recorded no oxygen saturation values. This shows that pulse oximeter is capable of detecting pulp vitality through enamel and dentin.
A Nellcor OxiMax' Dura Y D-YS multisite oxygen sensor was selected for use on teeth as the length and breadth of this sensor approximated with that of the tooth. It measured 4mm in length and 5mm in breadth. For holding these sensors parallel to the tooth, the dental holder was fabricated in such a way that it provided a stable relationship between the sensor elements and the tooth, thus maintaining a constant path length for the light between the emitter and detector required for pulse oximeter to make accurate calculations. This sensor holder along with the sensors is customized to be used on all teeth except lower anteriors due to their small size.
The mean oxygen saturation value for maxillary central incisor was 79.31 (SD=3.07), for maxillary lateral incisors 79.61 (SD=2.73) and for maxillary canines 79.85 (SD=2.09). These values correlate with the oxygen saturation values reported by Munshi AK and Hedge AM [2] where the mean oxygen saturation values for permanent central incisors was 81%, permanent right lateral incisors was 80.52%, permanent left lateral incisors was 80.75% and for index fingers used as control was 98.2%.
The lower oxygen saturation values for teeth compared to fingers can be attributed to several causes. Diffraction of the infrared light by the enamel prisms and dentin may cause decrease in oxygen saturation values. Fien et al suggested that the lower oxygen saturation values for the pulpal circulation may be attributed to light scatter through the gingiva [2].
Since reproducible oxygen saturation values are obtained using pulse oximeter, it has immediate clinical value by providing baseline clinical data for traumatized teeth. Further readings can then be compared to the baseline oxygen saturation values to determine any trends towards nonvitality. Further research, using a longitudinal study of pulse oximeter evaluation of traumatized teeth is indicated.
Conclusion | |  |
Consistent pulse oximeter readings in this study confirm that pulpal circulation and blood oxygen saturation can be detected by the customized dental pulse oximeter probe. This study shows that pulse oximetey is an effective, objective method of evaluating dental pulp vitality. In view of the limitations of the conventional pulp testing methods, pulse oximetry offers a more reliable and better evaluation of the changing patterns of circulation if any following trauma or other insults to the tooth.
Acknowledgement | |  |
1. This study was supported by the scientific research grant given by university of Meenakshi Academy of Higher Education and Research, Chennai, INDIA, vide Grant No.- MAIIER/742/2006.
2. Mr. Ilangovan, head of CAD-CAM center of Central Institute of Plastic Engineering and Technology, Chennai, INDIA for their collaborative support in the fabrication of custom made Pulse Oximeter Dental Sensor Holder.
3. Mr. A.K. Mathai, Biostatistician, Indian Council of Medical Reseach for statistical analysis
References | |  |
1. | Fratkin RD, Kenny DJ, J ohnston DIE Evaluation of a laser Doppler Flowmeter to assess blood flow in human primary incisor teeth, Pediatr Dent, 21:5356,1999. |
2. | Munshi A, Hedge AM, Radhakrishnan S: Pulse oximetry: a diagnostic instrument in pulp vitality testing, J Clin Pediatr Dent, 26:141-145,2002. |
3. | Noblett C Wilcox L, Scamman F: Detection of pulpal circulation in vitro by pulse oximetry, J Endod,22:1-5,1996. |
4. | Sasano T, Onodera D, Hashimoto K: Possible application of transmitted laser light for the assessment of human pulp vitality. Part 2. Increased laser power for enhanced detection of pulpal blood flow, Dent Traurnatol, 21:37-41,2005. |
5. | IkedaH.SudaH: Subjectivesensation andobjective neural discharges recorded from clinically nonvital and intactteeth, J Endod, 24:552-556,1998. |
6. | Yanpiset K, Vongsavan N, Sigurdsson A: Efficacy of laser doppler flowmetry for the diagnosis of revasculaization of reimplanted miniature dog teeth, Dent Traurnato1,17:63-7, 2001.. |
7. | Nisson R, Trope M, Zhang C.D: Dual wavelength spectrophotometry as a diagnostic test of the pulp chamber contents oral Sorg, 74:508-514,1992. |
8. | John W: Neonatal and pediatric Pulse Oximetey, Respiratory Care, 48:286-297,2003. |
9. | Schnettler JM, Wallace JA: Pulse oximeter as a diagnostic tool of pulp vitality, J Endod, 17:488490,1991. |
10. | Kahan RS, GulabivalaK, Snook M: Evaluation of a pulse oximeter and customized probe for pulp vitality testing, J Endod,22:105-109,1996. |

Correspondence Address: V Gopi Krishna Department of Conservative Dentistry and Endodontics, Meenakshi Ammal Dental College, Chennai, India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0970-9290.29880

Figures
[Figure - 1] Tables
[Table - 1] |
|
This article has been cited by | 1 |
Pulp Vitality Testing with a Developed Universal Pulse Oximeter Probe Holder |
|
| Živile Grabliauskiene, Roberta Zamaliauskiene, Greta Lodiene | | Medicina. 2021; 57(2): 101 | | [Pubmed] | [DOI] | | 2 |
Efficiency of new custom-made pulse oximeter sensor holder in assessment of actual pulp status |
|
| Krishnamachari Janani, P Ajitha, Raghu Sandhya, Haripriya Subbaiyan, Jerry Jose | | Journal of Family Medicine and Primary Care. 2020; 9(7): 3333 | | [Pubmed] | [DOI] | | 3 |
Interference of coronal enamel and dentin thickness and ambient light on pulse oximetry interpretation |
|
| Julio Almeida SILVA, Ana Helena Gonçalves de ALENCAR, Larissa Emanuelle SESTARI, Higor Almeida BARBOSA, Patrícia Correia de SIQUEIRA, Daniel Almeida DECURCIO, Fernando Branco BARLETTA, Carlos ESTRELA | | Brazilian Oral Research. 2020; 34 | | [Pubmed] | [DOI] | | 4 |
Reference values for pulp oxygen saturation as a diagnostic tool in endodontics: a systematic review and meta-analysis |
|
| Paula Lambert, Sergio Augusto Quevedo Miguens, Caroline Solda, Juliana Tomaz Sganzerla, Leandro Azambuja Reichert, Carlos Estrela, Fernando Branco Barletta | | Restorative Dentistry & Endodontics. 2020; 45(4) | | [Pubmed] | [DOI] | | 5 |
Thermographic assessment of vital and non-vital anterior teeth: A comparative study |
|
| Sara Mendes, Joaquim Mendes, André Moreira, Miguel Pais Clemente, Mário Vasconcelos | | Infrared Physics & Technology. 2020; 106: 103232 | | [Pubmed] | [DOI] | | 6 |
Oxygen Saturation in the Dental Pulp of Maxillary and Mandibular Molars - Part 2 |
|
| Carlos Estrela,Keila SA Oliveira,Ana Helena G Alencar,Fernando B Barletta,Cyntia RA Estrela,Wilson T Felippe | | Brazilian Dental Journal. 2017; 28(6): 704 | | [Pubmed] | [DOI] | | 7 |
Detection of Photoacoustic Signals from Blood in Dental Pulp |
|
| Azusa Yamada,Satoko Kakino,Yuji Matsuura | | Optics and Photonics Journal. 2016; 06(09): 229 | | [Pubmed] | [DOI] | | 8 |
Oxygen saturation and perfusion index from pulse oximetry in adult volunteers with viable incisors |
|
| Hyoun-Joong Kong,Teo Jeon Shin,Hong-Keun Hyun,Young-Jae Kim,Jung-Wook Kim,Won-Jun Shon | | Acta Odontologica Scandinavica. 2016; 74(5): 411 | | [Pubmed] | [DOI] | | 9 |
Oxygen saturation and perfusion index from pulse oximetry in adult volunteers with viable incisors |
|
| Hyoun-Joong Kong,Teo Jeon Shin,Hong-Keun Hyun,Young-Jae Kim,Jung-Wook Kim,Won-Jun Shon | | Acta Odontologica Scandinavica. 2016; 74(5): 411 | | [Pubmed] | [DOI] | | 10 |
The use of pulse oximetry in evaluation of pulp vitality in immature permanent teeth |
|
| Majid Bargrizan,Mohammad Asna Ashari,Mohsen Ahmadi,Jamileh Ramezani | | Dental Traumatology. 2016; 32(1): 43 | | [Pubmed] | [DOI] | | 11 |
The use of pulse oximetry in evaluation of pulp vitality in immature permanent teeth |
|
| Majid Bargrizan,Mohammad Asna Ashari,Mohsen Ahmadi,Jamileh Ramezani | | Dental Traumatology. 2016; 32(1): 43 | | [Pubmed] | [DOI] | | 12 |
Comparative Study of Pulp Vitality in Primary and Young Permanent Molars in Human Children with Pulse Oximeter and Electric Pulp Tester |
|
| Arun Sharma, Manish Madan, Prinka Shahi, PB Sood, Nishat Shahi | | International Journal of Clinical Pediatric Dentistry. 2015; 8(2): 94 | | [Pubmed] | [DOI] | | 13 |
Evaluation of pulp vitality tests in children: thermal test, electric pulp test and pulse oximetry |
|
| N Ahmadi,M Zare,N Kaviani,A Malekpour | | Iranian Journal of Pediatric Dentistry. 2014; 9(2): 31 | | [Pubmed] | [DOI] | | 14 |
Oxygen Saturation in the Dental Pulp of Permanent Teeth: A Critical Review |
|
| Kely Firmino Bruno,Fernando Branco Barletta,Wilson Tadeu Felippe,Júlio Almeida Silva,Ana Helena Gonçalves de Alencar,Carlos Estrela | | Journal of Endodontics. 2014; | | [Pubmed] | [DOI] | | 15 |
Clinical Diagnosis of Pulp Inflammation Based on Pulp Oxygenation Rates Measured by Pulse Oximetry |
|
| Frank C. Setzer,Simony Hidee Hamoy Kataoka,Fernanda Natrielli,Eudes Gondim-Junior,Celso Luiz Caldeira | | Journal of Endodontics. 2012; 38(7): 880 | | [Pubmed] | [DOI] | | 16 |
Comparison of the Efficacy of a Custom-made Pulse Oximeter Probe with Digital Electric Pulp Tester, Cold Spray, and Rubber Cup for Assessing Pulp Vitality |
|
| Nafiseh Dastmalchi,Hamid Jafarzadeh,Saeed Moradi | | Journal of Endodontics. 2012; 38(9): 1182 | | [Pubmed] | [DOI] | | 17 |
Comparison of the efficacy of a custom-made pulse oximeter probe with digital electric pulp tester, cold spray, and rubber cup for assessing pulp vitality |
|
| Dastmalchi, N. and Jafarzadeh, H. and Moradi, S. | | Journal of Endodontics. 2012; 38(9): 1182-1186 | | [Pubmed] | | 18 |
Clinical diagnosis of pulp inflammation based on pulp oxygenation rates measured by pulse oximetry |
|
| Setzer, F.C. and Kataoka, S.H.H. and Natrielli, F. and Gondim-Junior, E. and Caldeira, C.L. | | Journal of Endodontics. 2012; 38(7): 880-883 | | [Pubmed] | | 19 |
Comparison of the reliability of laser Doppler flowmetry, pulse oximetry and electric pulp tester in assessing the pulp vitality of human teeth |
|
| Karayilmaz, H. and Kirzioǧlu, Z. | | Journal of Oral Rehabilitation. 2011; 38(5): 340-347 | | [Pubmed] | | 20 |
Comparison of the reliability of laser Doppler flowmetry, pulse oximetry and electric pulp tester in assessing the pulp vitality of human teeth |
|
| H. KARAYILMAZ,Z. KIRZIOGLU | | Journal of Oral Rehabilitation. 2011; 38(5): 340 | | [Pubmed] | [DOI] | | 21 |
Laser Doppler flowmetry in endodontics |
|
| Jafarzadeh, H. | | International Endodontic Journal. 2009; 42(6): 476-490 | | [Pubmed] | | 22 |
Pulse oximetry: review of a potential aid in endodontic diagnosis |
|
| Jafarzadeh, H. and Rosenberg, P.A. | | Journal of Endodontics. 2009; 35(3): 329-333 | | [Pubmed] | | 23 |
Dental Pulp Neurophysiology: Part 2. Current Diagnostic Tests to Assess Pulp Vitality |
|
| Abd-Elmeguid, A. and BDS, M.D.S. and Yu, D.C. and CAGS, M.S.D. | | Journal (Canadian Dental Association). 2009; 75(2): 139-143 | | [Pubmed] | | 24 |
Assessment of pulp vitality: a review. |
|
| GOPIKRISHNA, V. and PRADEEP, G. and VENKATESHBABU, N. | | International Journal of Paediatric Dentistry. 2009; 19(1): 3-15 | | [Pubmed] | | 25 |
Pulse Oximetry: Review of a Potential Aid in Endodontic Diagnosis |
|
| Hamid Jafarzadeh,Paul A. Rosenberg | | Journal of Endodontics. 2009; 35(3): 329 | | [Pubmed] | [DOI] | | 26 |
Laser Doppler flowmetry in endodontics: a review |
|
| H. Jafarzadeh | | International Endodontic Journal. 2009; 42(6): 476 | | [Pubmed] | [DOI] | | 27 |
Assessment of pulp vitality: a review |
|
| VELAYUTHAM GOPIKRISHNA,GALI PRADEEP,NAGENDRABABU VENKATESHBABU | | International Journal of Paediatric Dentistry. 2009; 19(1): 3 | | [Pubmed] | [DOI] | | 28 |
Evaluation of Efficacy of a New Custom-Made Pulse Oximeter Dental Probe in Comparison With the Electrical and Thermal Tests for Assessing Pulp Vitality |
|
| Velayutham Gopikrishna,Kush Tinagupta,Deivanayagam Kandaswamy | | Journal of Endodontics. 2007; 33(4): 411 | | [Pubmed] | [DOI] | | 29 |
Comparison of Electrical, Thermal, and Pulse Oximetry Methods for Assessing Pulp Vitality in Recently Traumatized Teeth |
|
| Velayutham Gopikrishna,Kush Tinagupta,Deivanayagam Kandaswamy | | Journal of Endodontics. 2007; 33(5): 531 | | [Pubmed] | [DOI] | | 30 |
Comparison of Electrical, Thermal, and Pulse Oximetry Methods for Assessing Pulp Vitality in Recently Traumatized Teeth |
|
| Gopikrishna, V., Tinagupta, K., Kandaswamy, D. | | Journal of Endodontics. 2007; 33(5): 531-535 | | [Pubmed] | | 31 |
Evaluation of Efficacy of a New Custom-Made Pulse Oximeter Dental Probe in Comparison With the Electrical and Thermal Tests for Assessing Pulp Vitality |
|
| Gopikrishna, V., Tinagupta, K., Kandaswamy, D. | | Journal of Endodontics. 2007; 33(4): 411-414 | | [Pubmed] | |
|
|
 |
 |
|
|
|
|
|
|
Article Access Statistics | | Viewed | 13532 | | Printed | 576 | | Emailed | 9 | | PDF Downloaded | 1092 | | Comments | [Add] | | Cited by others | 31 | |
|

|